CVE-2019-25720 json

Dräger SC Monitoring Devices DoS via Malformed Network Packet

published 2026-06-03 5 mo ago modified 2026-07-22 3 mo ago also EUVD-2019-20161

cvss
6.5
version 3.1, CNA
cisa kev
Not listed
not in CISA's catalog
epss
0.2%
EPSS 0.2% chance of exploitation in 30 days, percentile 9%
public exploits
0
none known

Summary

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.

CISA SSVC

Exploitation
none
Automatable
no
Technical impact
partial
Assessed
5 mo ago

Exploit prediction

0.20% chance of exploitation in the next 30 days, higher than 9% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
CNA 4.0 7.1 CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N
CNA 3.1 6.5 CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
CNA 4.0 7.1 CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

Affected products

  • Dräger SC 6002XL
  • Dräger SC6802XL
  • Dräger SC 7000
  • Dräger SC8000
  • Dräger SC90000 XL

Classification

Weakness
  • CWE-1286 Improper Validation of Syntactic Correctness of Input