Skip to content
Fencer Vulns
high 8.2

CVE-2018-10601 json

IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550)…

published 2018-06-05 8 y ago modified 2026-06-17 4 mo ago also EUVD-2018-2673

cvss
8.2
version 3.1, NVD
cisa kev
Not listed
not in CISA's catalog
epss
0.4%
EPSS 0.4% chance of exploitation in 30 days, percentile 31%
public exploits
0
none known

Summary

IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-18-156-01 updated cvss 8.2

Philips' IntelliVue Patient and Avalon Fetal Monitors

  • Philips Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 F.0.G.0, J.3 Philips IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M Philips IntelliVue Patient Monitors MX (MX400-550) Rev J-M, (X3/MX100 for Rev M only)
    affected: vendor fix philips.com

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

Exploit prediction

0.37% chance of exploitation in the next 30 days, higher than 31% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
NVD 3.1 8.2 CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H
NVD 2.0 5.4 AV:A/AC:M/Au:N/C:P/I:P/A:P

Affected products

  • ICS-CERT IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors

Classification

Weakness

References

Sources

Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.