medium 5.3

CVE-2018-10599 json

IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550)…

published 2018-06-05 8 y ago modified 2026-06-17 4 mo ago also EUVD-2018-2671

cvss
5.3
version 3.0, NVD
cisa kev
Not listed
not in CISA's catalog
epss
0.4%
EPSS 0.4% chance of exploitation in 30 days, percentile 36%
public exploits
0
none known

Summary

IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that allows an unauthenticated attacker to read memory from an attacker-chosen device address within the same subnet.

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-18-156-01 updated cvss 6.4

Philips' IntelliVue Patient and Avalon Fetal Monitors

  • Philips Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 F.0.G.0, J.3 Philips IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M Philips IntelliVue Patient Monitors MX (MX400-550) Rev J-M, (X3/MX100 for Rev M only)
    affected: vendor fix philips.com

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

Exploit prediction

0.42% chance of exploitation in the next 30 days, higher than 36% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
NVD 3.0 5.3 CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
NVD 2.0 2.9 AV:A/AC:M/Au:N/C:P/I:N/A:N

Affected products

  • ICS-CERT IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors

Classification

Weakness
  • CWE-200 Exposure of Sensitive Information to an Unauthorized Actor

References

Sources

Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.