CVE-2017-12714 json
Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" c…
published 2018-04-25 8 y ago modified 2026-06-17 4 mo ago also EUVD-2017-4253
- cvss
- 6.5
- version 3.0, NVD
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.6%
- EPSS 0.6% chance of exploitation in 30 days, percentile 50%
- public exploits
- 0
- none known
Summary
Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Patch / fix
- abbott accent_firmware f0b.0e.7e
- abbott anthem_firmware f0b.0e.7e
- abbott accent_mri_firmware f10.08.6c
- abbott accent_st_firmware f10.08.6c
- abbott assurity_firmware f14.07.80
- abbott allure_firmware f14.07.80
- abbott assurity_mri_firmware f17.01.49
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-18-107-01 updated cvss 5.3
Abbott Laboratories Defibrillator
-
Abbott Laboratories Current distributed < april 1, 2018 Abbott Laboratories Ellipse distributed < april 1, 2018 Abbott Laboratories Fortify distributed < april 1, 2018 Abbott Laboratories Fortify Assura distributed < april 1, 2018 Abbott Laboratories Promote distributed < april 1, 2018 Abbott Laboratories Promote Quadra distributed < april 1, 2018affected: vendor fix sjm.com
+5 more productsfewer
Abbott Laboratories Quadra Assura distributed < april 1, 2018 Abbott Laboratories Quadra Assura MP distributed < april 1, 2018 Abbott Laboratories Unify distributed < april 1, 2018 Abbott Laboratories Unify Assura distributed < april 1, 2018 Abbott Laboratories Unify Quadra distributed < april 1, 2018
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
cisa ICSMA-17-241-01 updated cvss 5.3
ICSMA-17-241-01_Abbott Laboratories ' Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI Pacemaker Vulnerabilities
-
Abbott Laboratories Accent MRI < august 28 Abbott Laboratories Accent/Anthem < august 28 Abbott Laboratories Assurity MRI < august 28 Abbott Laboratories Assurity/Allure < august 28affected: mitigation
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.63% chance of exploitation in the next 30 days, higher than 50% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| NVD | 3.0 | 6.5 | CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H |
| NVD | 2.0 | 6.1 | AV:A/AC:L/Au:N/C:N/I:N/A:C |
Affected products
- Abbott Laboratories Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI.
Classification
- Weakness
-
- CWE-920 Improper Restriction of Power Consumption
References
-
ics-cert.us-cert.gov
/advisories/ICSMA-17-241-01
third-party-advisory, us-government-resource
-
securityfocus.com
/bid/100523
vdb-entry, third-party-advisory
Sources
- CVE List CVE-2017-12714
- NVD CVE-2017-12714