CVE-2017-12701 json
BMC Medical Luna CPAP Machines released prior to July 1, 2017, contain an improper input validation vulnerability which may allow an authen…
published 2018-04-17 8 y ago modified 2026-06-17 4 mo ago also EUVD-2017-4240
- cvss
- 6.5
- version 3.0, NVD
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 1.3%
- EPSS 1.3% chance of exploitation in 30 days, percentile 70%
- public exploits
- 0
- none known
Summary
BMC Medical Luna CPAP Machines released prior to July 1, 2017, contain an improper input validation vulnerability which may allow an authenticated attacker to crash the CPAP's Wi-Fi module resulting in a denial-of-service condition.
Patch / fix
- cpap luna_cpap_machine_firmware 20170701
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-17-227-01 updated cvss 4.6
ICSMA-17-227-01_BMC Medical and 3B Medical Luna CPAP Machine
-
BMC Medical, 3B Medical Luna CPAP Machine < july 1 2017affected: mitigation “This vulnerability has been addressed in devices released after July 1, 2017. For devices released prior to July 1, 2017, BMC Medical and 3B Medical offer no mitigations.”
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
1.28% chance of exploitation in the next 30 days, higher than 70% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| NVD | 3.0 | 6.5 | CVSS:3.0/AV:N/AC:L/PR:L/UI:N/S:U/C:N/I:N/A:H |
| NVD | 2.0 | 4.0 | AV:N/AC:L/Au:S/C:N/I:N/A:P |
Affected products
- BMC Medical Luna CPAP Machine
Classification
- Weakness
-
- CWE-20 Improper Input Validation
References
-
securityfocus.com
/bid/100354
vdb-entry, third-party-advisory
-
ics-cert.us-cert.gov
/advisories/ICSMA-17-227-01
third-party-advisory, us-government-resource
Sources
- CVE List CVE-2017-12701
- NVD CVE-2017-12701
Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.