CVE-2017-11580 json

Blipcare Wifi blood pressure monitor BP700 10.1 devices allow memory corruption that results in Denial of Service.

published 2019-07-02 7 y ago modified 2026-06-17 4 mo ago also EUVD-2017-3195

cvss
6.5
version 3.0, NVD
cisa kev
Not listed
not in CISA's catalog
epss
1.4%
EPSS 1.4% chance of exploitation in 30 days, percentile 73%
public exploits
0
none known

Summary

Blipcare Wifi blood pressure monitor BP700 10.1 devices allow memory corruption that results in Denial of Service. When connected to the "Blip" open wireless connection provided by the device, if a large string is sent as a part of the HTTP request in any part of the HTTP headers, the device could become completely unresponsive. Presumably this happens as the memory footprint provided to this device is very small. According to the specs from Rezolt, the Wi-Fi module only has 256k of memory. As a result, an incorrect string copy operation using either memcpy, strcpy, or any of their other variants could result in filling up the memory space allocated to the function executing and this would result in memory corruption. To test the theory, one can modify the demo application provided by the Cypress WICED SDK and introduce an incorrect "memcpy" operation and use the compiled application on the evaluation board provided by Cypress semiconductors with exactly the same Wi-Fi SOC. The results were identical where the device would completely stop responding to any of the ping or web requests.

Exploit prediction

1.43% chance of exploitation in the next 30 days, higher than 73% of all scored CVEs (FIRST EPSS, 2026-11-08).

Exploits and checks

  • github.com https://github.com/ethanhunnt/IoT_vulnerabilities/blob/master/Blipcare_sec_issues.pdf referenced PoC, tagged "exploit" by a source, unverified

CVSS scores

assigned by version score vector
NVD 3.0 6.5 CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H