CVE-2018-10601 json
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550)…
published 2018-06-05 8 y ago modified 2026-06-17 4 mo ago also EUVD-2018-2673
- cvss
- 8.2
- version 3.1, NVD
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.4%
- EPSS 0.4% chance of exploitation in 30 days, percentile 31%
- public exploits
- 0
- none known
Summary
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-18-156-01 updated cvss 8.2
Philips' IntelliVue Patient and Avalon Fetal Monitors
-
Philips Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 F.0.G.0, J.3 Philips IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M Philips IntelliVue Patient Monitors MX (MX400-550) Rev J-M, (X3/MX100 for Rev M only)affected: vendor fix philips.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.37% chance of exploitation in the next 30 days, higher than 31% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| NVD | 3.1 | 8.2 | CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H |
| NVD | 2.0 | 5.4 | AV:A/AC:M/Au:N/C:P/I:P/A:P |
Affected products
- ICS-CERT IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors
Classification
References
-
ics-cert.us-cert.gov
/advisories/ICSMA-18-156-01
third-party-advisory, us-government-resource
Sources
- CVE List CVE-2018-10601
- NVD CVE-2018-10601