Skip to content
Fencer Vulns
medium 5.9

CVE-2025-0683 json

Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor

published 2025-01-30 21 mo ago modified 2026-06-17 4 mo ago also EUVD-2025-1815

cvss
5.9
version 3.1, CNA
cisa kev
Not listed
not in CISA's catalog
epss
0.8%
EPSS 0.8% chance of exploitation in 30 days, percentile 56%
public exploits
0
none known

Summary

In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-25-030-01 updated cvss 5.9

Contec Health CMS8000 Patient Monitor (Update A)

  • Contec Health CMS8000 Patient Monitor Contec Health CMS8000 Patient Monitor Firmware smart3250-2.6.27-wlan2.1.7.cramfs, CMS7.820.075.08/0.74(0.75), CMS7.820.120.01/0.93(0.95)
    affected: mitigation “Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.” fda.gov

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

CISA SSVC

Exploitation
none
Automatable
no
Technical impact
partial
Assessed
21 mo ago

Exploit prediction

0.78% chance of exploitation in the next 30 days, higher than 56% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
CNA 4.0 8.2 CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N
CNA 3.1 5.9 CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
CNA 4.0 8.2 CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

Affected products

  • Contec Health CMS8000 Patient Monitor

Classification

Weakness
  • CWE-359 Exposure of Private Personal Information to an Unauthorized Actor