Skip to content
Fencer Vulns
high 7.6

CVE-2019-25722 json

Dräger SC Monitoring Devices Hard-coded Credentials and DoS

published 2026-06-02 5 mo ago modified 2026-07-22 3 mo ago also EUVD-2019-20158

cvss
7.6
version 3.1, CNA
cisa kev
Not listed
not in CISA's catalog
epss
0.2%
EPSS 0.2% chance of exploitation in 30 days, percentile 9%
public exploits
0
none known

Summary

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.

CISA SSVC

Exploitation
none
Automatable
no
Technical impact
partial
Assessed
5 mo ago

Exploit prediction

0.19% chance of exploitation in the next 30 days, higher than 9% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
CNA 4.0 7.2 CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N
CNA 3.1 7.6 CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
CNA 4.0 7.2 CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

Affected products

  • Dräger SC 6002XL
  • Dräger SC6802XL
  • Dräger SC 7000
  • Dräger SC8000
  • Dräger SC90000 XL

Classification

Weakness
  • CWE-798 Use of Hard-coded Credentials

References

Sources

Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.