Skip to content
Fencer Vulns
medium 6.5

CVE-2017-9658 json

Certain 802.11 network management messages have been determined to invoke wireless access point blacklisting security defenses when not req…

published 2018-04-30 8 y ago modified 2026-06-17 4 mo ago also EUVD-2017-18589

cvss
6.5
version 3.0, NVD
cisa kev
Not listed
not in CISA's catalog
epss
0.8%
EPSS 0.8% chance of exploitation in 30 days, percentile 57%
public exploits
0
none known

Summary

Certain 802.11 network management messages have been determined to invoke wireless access point blacklisting security defenses when not required, which can necessitate intervention by hospital staff to reset the device and reestablish a network connection to the Wi-Fi access point. During this state, the Philips IntelliVue MX40 Version B.06.18 can either connect to an alternative access point within signal range for association to a central monitoring station, or it can remain in local monitoring mode until the device is reset by hospital staff. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

Patch / fix

  • philips intellivue_mx40_firmware b.06.18

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-17-255-01 updated cvss 6.5

ICSMA-17-255-01_Philips' IntelliView MX40 Patient Worn Monitor (WLAN) Vulnerabilities

  • Philips IntelliVue MX40 Patient Worn Monitor (WLAN only) < B.06.18
    affected: mitigation usa.philips.com

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

Exploit prediction

0.81% chance of exploitation in the next 30 days, higher than 57% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
NVD 3.0 6.5 CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
NVD 2.0 6.1 AV:A/AC:L/Au:N/C:N/I:N/A:C

Affected products

  • Philips IntelliVue MX40 Patient Worn Monitor

Classification

Weakness
  • CWE-755 Improper Handling of Exceptional Conditions

References

Sources

Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.