CVE-2026-66832 json
Mira Hormone Monitor, Mira Android App Use of GET request method with sensitive query strings
published 2026-08-11 2 mo ago modified 2026-09-01 2 mo ago also EUVD-2026-56978
- cvss
- 6.5
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.4%
- EPSS 0.4% chance of exploitation in 30 days, percentile 33%
- public exploits
- 0
- none known
Summary
When the Mira Android app opens in-app WebView content (e.g., shop redirect flows), the user's live session token is appended to the URL as a query string parameter, and a persistent user identifier is included in the WebView's User-Agent header. Both are then transmitted to third-party web properties, referrer logs, and any JavaScript running in the WebView context.
Patch / fix
- Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Firmware 01.07.01.53
- Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Android App 4.5.18
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-26-223-01 updated cvss 6.5
Mira Hormone Monitor, Mira Android App
-
Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Android App 4.5.15.4 Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Monitor Firmware 1.7.1.47affected: mitigation “Users should update the Mira app to the latest version iOS v3.5.18 / Android v4.5.18. Firmware v01.07.01.53 is updated via the app when the device is connected. No additional action is required.”
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
CISA SSVC
- Exploitation
- none
- Automatable
- no
- Technical impact
- partial
- Assessed
- 2 mo ago
Exploit prediction
0.39% chance of exploitation in the next 30 days, higher than 33% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 4.0 | 6.9 | CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:A/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N |
| CNA | 3.1 | 6.5 | CVSS:3.1/AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:N/A:N |
| CNA | 4.0 | 6.9 | CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:A/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X |
Affected products
- Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Firmware
- Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Android App
Classification
- Weakness
-
- CWE-598 Use of HTTP Request With Sensitive Query String
References
Sources
- CVE List CVE-2026-66832
- NVD CVE-2026-66832
Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.