CVE-2025-24836 json
Qardio Heart Health IOS and Android Application and QardioARM A100 Uncaught Exception
published 2025-02-13 20 mo ago modified 2026-06-17 4 mo ago also EUVD-2025-3959
- cvss
- 7.1
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.2%
- EPSS 0.2% chance of exploitation in 30 days, percentile 15%
- public exploits
- 0
- none known
Summary
With a specially crafted Python script, an attacker could send continuous startMeasurement commands over an unencrypted Bluetooth connection to the affected device. This would prevent the device from connecting to a clinician's app to take patient readings and ostensibly flood it with requests, resulting in a denial-of-service condition.
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-25-044-01 updated cvss 7.1
Qardio Heart Health IOS and Android Application and QardioARM A100
-
Qardio Heart Health Android Mobile Application 2.5.1 Qardio Heart Health IOS Mobile Application 2.7.4 QardioARM A100affected: mitigation “Qardio has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Qardio customer support for additional information.” qardio.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
CISA SSVC
- Exploitation
- none
- Automatable
- no
- Technical impact
- total
- Assessed
- 20 mo ago
Exploit prediction
0.24% chance of exploitation in the next 30 days, higher than 15% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.1 | 7.1 | CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H |
| CNA | 4.0 | 6.1 | CVSS:4.0/AV:A/AC:H/AT:N/PR:N/UI:N/VC:L/VI:H/VA:H/SC:N/SI:N/SA:N |
| CNA | 4.0 | 6.1 | CVSS:4.0/AV:A/AC:H/AT:N/PR:N/UI:N/VC:L/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X |
Affected products
- Qardio Heart Health IOS Mobile Application
- Qardio Heart Health Android Mobile Application
- Qardio QardioARM
Classification
- Weakness
-
- CWE-248 Uncaught Exception
References
Sources
- CVE List CVE-2025-24836
- NVD CVE-2025-24836
Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.