Skip to content
Fencer Vulns
high 7.1

CVE-2025-24836 json

Qardio Heart Health IOS and Android Application and QardioARM A100 Uncaught Exception

published 2025-02-13 20 mo ago modified 2026-06-17 4 mo ago also EUVD-2025-3959

cvss
7.1
version 3.1, CNA
cisa kev
Not listed
not in CISA's catalog
epss
0.2%
EPSS 0.2% chance of exploitation in 30 days, percentile 15%
public exploits
0
none known

Summary

With a specially crafted Python script, an attacker could send continuous startMeasurement commands over an unencrypted Bluetooth connection to the affected device. This would prevent the device from connecting to a clinician's app to take patient readings and ostensibly flood it with requests, resulting in a denial-of-service condition.

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-25-044-01 updated cvss 7.1

Qardio Heart Health IOS and Android Application and QardioARM A100

  • Qardio Heart Health Android Mobile Application 2.5.1 Qardio Heart Health IOS Mobile Application 2.7.4 QardioARM A100
    affected: mitigation “Qardio has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Qardio customer support for additional information.” qardio.com

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

CISA SSVC

Exploitation
none
Automatable
no
Technical impact
total
Assessed
20 mo ago

Exploit prediction

0.24% chance of exploitation in the next 30 days, higher than 15% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
CNA 3.1 7.1 CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H
CNA 4.0 6.1 CVSS:4.0/AV:A/AC:H/AT:N/PR:N/UI:N/VC:L/VI:H/VA:H/SC:N/SI:N/SA:N
CNA 4.0 6.1 CVSS:4.0/AV:A/AC:H/AT:N/PR:N/UI:N/VC:L/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

Affected products

  • Qardio Heart Health IOS Mobile Application
  • Qardio Heart Health Android Mobile Application
  • Qardio QardioARM

Classification

Weakness

References

Sources

Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.

CPEs and fixed versions use Anchore's correction of NVD's data: Added CPE configurations because not yet analyzed by NVD.