CVE-2025-14097 json
Remote Code Execution Vulnerability in Radiometer Products
published 2025-12-17 10 mo ago modified 2026-06-17 4 mo ago also EUVD-2025-203895
- cvss
- 7.2
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.5%
- EPSS 0.5% chance of exploitation in 30 days, percentile 40%
- public exploits
- 0
- none known
Summary
A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software. Other related CVE's are CVE-2025-14095 & CVE-2025-14096. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer. Temporary work Around: If the network is not considered secure, please remove the analyzer from the network. Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status:
Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.
Patch / fix
- Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Application software versions >= 3.5MR11 with Windows 10 operating system
- Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers Application software versions >= 6.20MR2 with Windows 10 operating system
- Radiometer Medical Aps AQT90 FLEX Analyzers Application software versions >= 8.13 MR2 with Windows 10 operating system
CISA SSVC
- Exploitation
- none
- Automatable
- no
- Technical impact
- total
- Assessed
- 10 mo ago
Exploit prediction
0.46% chance of exploitation in the next 30 days, higher than 40% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.1 | 7.2 | CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H |
Affected products
- Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers
- Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers
- Radiometer Medical Aps AQT90 FLEX Analyzers
Classification
- Weakness
-
- CWE-287 Improper Authentication
References
Sources
- CVE List CVE-2025-14097
- NVD CVE-2025-14097
Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.