Skip to content
Fencer Vulns
medium 6.8

CVE-2025-14095 json

Privilege boundary violation in Radiometer Products

published 2025-12-17 10 mo ago modified 2026-10-07 33 d ago also EUVD-2025-203887

cvss
6.8
version 3.1, CNA
cisa kev
Not listed
not in CISA's catalog
epss
0.2%
EPSS 0.2% chance of exploitation in 30 days, percentile 5%
public exploits
0
none known

Summary

A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software.

Other related CVE's are CVE-2025-14096 & CVE-2025-14097.

Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.

Required configuration for Exposure:

Physical access to the analyzer is needed.

Temporary work Around:

Only authorized people can physically access the analyzer.

Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution.

Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication.                                                                                                                                                                                        Note:

CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.

CISA SSVC

Exploitation
none
Automatable
no
Technical impact
partial
Assessed
10 mo ago

Exploit prediction

0.17% chance of exploitation in the next 30 days, higher than 5% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
CNA 3.1 6.8 CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CNA 3.1 5.7 CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H

Affected products

  • Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers
  • Radiometer Medical Aps AQT90 FLEX Analyzers
  • Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers
  • Radiometer Medical Aps ABL9 Analyzers

Classification

Weakness
  • CWE-284 Improper Access Control
  • CWE-693 Protection Mechanism Failure

References

Sources

Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.