CVE-2025-1204 json
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existin…
published 2025-02-25 20 mo ago modified 2026-06-17 4 mo ago also EUVD-2025-5402
- cvss
- 7.7
- version 4.0, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.5%
- EPSS 0.5% chance of exploitation in 30 days, percentile 41%
- public exploits
- 0
- none known
Summary
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-25-030-01 updated cvss 7.5
Contec Health CMS8000 Patient Monitor (Update A)
-
Contec Health CMS8000 Patient Monitor Contec Health CMS8000 Patient Monitor Firmware smart3250-2.6.27-wlan2.1.7.cramfs, CMS7.820.075.08/0.74(0.75), CMS7.820.120.01/0.93(0.95)affected: mitigation “Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.” fda.gov
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
CISA SSVC
- Exploitation
- Proof of concept is public
- Automatable
- no
- Technical impact
- total
- Assessed
- 20 mo ago
Exploit prediction
0.48% chance of exploitation in the next 30 days, higher than 41% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 4.0 | 7.7 | CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N |
| CNA | 4.0 | 7.7 | CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X |
Affected products
- Contec Health CMS8000 Patient Monitor
Classification
- Weakness
-
- CWE-912 Hidden Functionality
References
Sources
- CVE List CVE-2025-1204
- NVD CVE-2025-1204