high 7.7

CVE-2025-1204 json

The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existin…

published 2025-02-25 20 mo ago modified 2026-06-17 4 mo ago also EUVD-2025-5402

cvss
7.7
version 4.0, CNA
cisa kev
Not listed
not in CISA's catalog
epss
0.5%
EPSS 0.5% chance of exploitation in 30 days, percentile 41%
public exploits
0
none known

Summary

The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-25-030-01 updated cvss 7.5

Contec Health CMS8000 Patient Monitor (Update A)

  • Contec Health CMS8000 Patient Monitor Contec Health CMS8000 Patient Monitor Firmware smart3250-2.6.27-wlan2.1.7.cramfs, CMS7.820.075.08/0.74(0.75), CMS7.820.120.01/0.93(0.95)
    affected: mitigation “Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.” fda.gov

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

CISA SSVC

Exploitation
Proof of concept is public
Automatable
no
Technical impact
total
Assessed
20 mo ago

Exploit prediction

0.48% chance of exploitation in the next 30 days, higher than 41% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
CNA 4.0 7.7 CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
CNA 4.0 7.7 CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

Affected products

  • Contec Health CMS8000 Patient Monitor

Classification

Weakness

References