CVE-2025-0626 json

Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor

published 2025-01-30 21 mo ago modified 2026-06-17 4 mo ago also EUVD-2025-1797

cvss
7.5
version 3.1, CNA
cisa kev
Not listed
not in CISA's catalog
epss
1.1%
EPSS 1.1% chance of exploitation in 30 days, percentile 66%
public exploits
0
none known

Summary

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-25-030-01 updated cvss 7.5

Contec Health CMS8000 Patient Monitor (Update A)

  • Contec Health CMS8000 Patient Monitor Contec Health CMS8000 Patient Monitor Firmware smart3250-2.6.27-wlan2.1.7.cramfs, CMS7.820.075.08/0.74(0.75), CMS7.820.120.01/0.93(0.95)
    affected: mitigation “Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.” fda.gov

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

CISA SSVC

Exploitation
none
Automatable
no
Technical impact
total
Assessed
21 mo ago

Exploit prediction

1.14% chance of exploitation in the next 30 days, higher than 66% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
CNA 4.0 7.7 CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
CNA 3.1 7.5 CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H
CNA 4.0 7.7 CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

Affected products

  • Contec Health CMS8000 Patient Monitor