CVE-2024-9834 json
Improper data protection on Life2000 ventilator serial interface
published 2024-11-14 23 mo ago modified 2026-06-17 4 mo ago also EUVD-2024-50149
- cvss
- 9.3
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.1%
- EPSS 0.1% chance of exploitation in 30 days, percentile 3%
- public exploits
- 0
- none known
Summary
Improper data protection on the ventilator's serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-24-319-01 updated cvss 9.3
Baxter Life2000 Ventilation System
-
Baxter Life2000 Ventilation System <=06.08.00.00affected: vendor fix, mitigation “Baxter plans to issue a follow-up announcement in Q2 2025 regarding the Life2000 vulnerabilities described in this disclosure.” baxter.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
CISA SSVC
- Exploitation
- none
- Automatable
- no
- Technical impact
- total
- Assessed
- 23 mo ago
Exploit prediction
0.14% chance of exploitation in the next 30 days, higher than 3% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.1 | 9.3 | CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
Affected products
- Baxter Life2000 Ventilation System
- baxter life2000_ventilator_firmware
Classification
- Weakness
-
- CWE-319 Cleartext Transmission of Sensitive Information
References