CVE-2023-25931 json
Medtronic Micro Clinician & InterStim X Clinician App Password Reset Issue
published 2023-03-01 3 y ago modified 2026-06-17 4 mo ago also EUVD-2023-29819
- cvss
- 6.4
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.3%
- EPSS 0.3% chance of exploitation in 30 days, percentile 18%
- public exploits
- 0
- none known
Summary
Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.
Patch / fix
- Medtronic InsterStim Applications A51200, A51300
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-23-061-01 updated cvss 6.4
Medtronic Micro Clinician and InterStim Apps
-
Medtronic InterStim X Clinician A51300 Medtronic Micro Clinician A51200affected: vendor fix global.medtronic.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
CISA SSVC
- Exploitation
- none
- Automatable
- no
- Technical impact
- partial
- Assessed
- 20 mo ago
Exploit prediction
0.27% chance of exploitation in the next 30 days, higher than 18% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.1 | 6.4 | CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H |
| NVD | 3.1 | 6.8 | CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
Affected products
- Medtronic InsterStim Applications
Classification
References
Sources
- CVE List CVE-2023-25931
- NVD CVE-2023-25931
Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.