CVE-2023-1966 json
CVE-2023-1966
published 2023-04-28 3 y ago modified 2026-06-17 4 mo ago also EUVD-2023-24152
- cvss
- 7.4
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.9%
- EPSS 0.9% chance of exploitation in 30 days, percentile 60%
- public exploits
- 0
- none known
Summary
Instruments with Illumina Universal Copy Service v1.x and v2.x contain an unnecessary privileges vulnerability. An unauthenticated malicious actor could upload and execute code remotely at the operating system level, which could allow an attacker to change settings, configurations, software, or access sensitive data on the affected product.
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-23-117-01 updated cvss 7.4
Illumina Universal Copy Service
-
Illumina iScan Control Software 4.0.0, 4.0.5 Illumina iSeq 100 Illumina MiniSeq Control Software >= 2.0 Illumina MiSeq Control Software 4.0 Illumina MiSeqDx Operating Software >= 4.0.1 Illumina NextSeq 500/550 Control Software 4.0affected: vendor fix support.illumina.com
+5 more productsfewer
Illumina NextSeq 550Dx Control Software 4.0 Illumina NextSeq 550Dx Operating Software >= 1.0.0, <= 1.3.1, >= 1.3.3 Illumina NextSeq 1000/2000 Control Software <= 1.4.1 Illumina NovaSeq 6000 Control Software <= 1.7 Illumina NovaSeq Control Software 1.8
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
CISA SSVC
- Exploitation
- none
- Automatable
- no
- Technical impact
- total
- Assessed
- 21 mo ago
Exploit prediction
0.92% chance of exploitation in the next 30 days, higher than 60% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.1 | 7.4 | CVSS:3.1/AV:P/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H |
| NVD | 3.1 | 9.8 | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
Affected products
- Illumina iScan Control Software
- Illumina MiniSeq Control Software
- Illumina iSeq 100
- Illumina MiSeq Control Software
- Illumina MiSeqDx Operating Software
- Illumina NextSeq 500/550 Control Software
- Illumina NextSeq 550Dx Control Software
- Illumina NextSeq 550Dx Operating Software
- Illumina NextSeq 1000/2000 Control Software
- Illumina NovaSeq 6000 Control Software
- Illumina NovaSeq Control Software
Classification
References
-
support.illumina.com
/downloads/illumina-universal-copy-service-1-0.html
vendor-advisory
-
cisa.gov
/news-events/ics-medical-advisories/icsma-23-117-01
third-party-advisory, us-government-resource
Sources
- CVE List CVE-2023-1966
- NVD CVE-2023-1966
Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.