CVE-2022-41627 json

The physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-…

published 2022-10-27 4 y ago modified 2026-06-17 4 mo ago also EUVD-2022-44818

cvss
4.8
version 3.1, CNA
cisa kev
Not listed
not in CISA's catalog
epss
0.1%
EPSS 0.1% chance of exploitation in 30 days, percentile 1%
public exploits
0
none known

Summary

The physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone’s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-22-298-01 updated cvss 4.8

AliveCor KardiaMobile

  • AliveCor KardiaMobile IoT device
    affected: mitigation “AliveCor has acknowledged that the data-over-sound protocol has no encryption and is vulnerable to CVE-2022-41627, but the circumstances necessary for exploitation are "unlikely."”

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

CISA SSVC

Exploitation
none
Automatable
no
Technical impact
partial
Assessed
18 mo ago

Exploit prediction

0.12% chance of exploitation in the next 30 days, higher than 1% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
CNA 3.1 4.8 CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:L/A:H
NVD 3.1 7.6 CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H

Affected products

  • AliveCor KardiaMobile

Classification

Weakness
  • CWE-311 Missing Encryption of Sensitive Data
  • CWE-319 Cleartext Transmission of Sensitive Information

References

Sources

Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.