Skip to content
Fencer Vulns
high 7.5

CVE-2022-38100 json

Contec Health CMS8000

published 2022-09-13 4 y ago modified 2026-06-17 4 mo ago also EUVD-2022-40702

cvss
7.5
version 3.1, CNA
cisa kev
Not listed
not in CISA's catalog
epss
1.0%
EPSS 1.0% chance of exploitation in 30 days, percentile 64%
public exploits
0
none known

Summary

The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-22-244-01 updated cvss 7.5

Contec Health CMS8000 Patient Monitor (Update A)

  • Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor <=smart3250-2.6.27-wlan2.1.7.cramfs, <=CMS7.820.075.08/0.740.75, <=CMS7.820.120.01/0.930.95
    affected: mitigation “Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.”

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

CISA SSVC

Exploitation
none
Automatable
yes
Technical impact
partial
Assessed
18 mo ago

Exploit prediction

1.05% chance of exploitation in the next 30 days, higher than 64% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
CNA 3.1 7.5 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
NVD 3.1 7.5 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H

Affected products

  • Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor

Classification

Weakness
  • CWE-400 Uncontrolled Resource Consumption

References

Sources

Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.