CVE-2022-38100 json
Contec Health CMS8000
published 2022-09-13 4 y ago modified 2026-06-17 4 mo ago also EUVD-2022-40702
- cvss
- 7.5
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 1.0%
- EPSS 1.0% chance of exploitation in 30 days, percentile 64%
- public exploits
- 0
- none known
Summary
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-22-244-01 updated cvss 7.5
Contec Health CMS8000 Patient Monitor (Update A)
-
Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor <=smart3250-2.6.27-wlan2.1.7.cramfs, <=CMS7.820.075.08/0.740.75, <=CMS7.820.120.01/0.930.95affected: mitigation “Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.”
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
CISA SSVC
- Exploitation
- none
- Automatable
- yes
- Technical impact
- partial
- Assessed
- 18 mo ago
Exploit prediction
1.05% chance of exploitation in the next 30 days, higher than 64% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.1 | 7.5 | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H |
| NVD | 3.1 | 7.5 | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H |
Affected products
- Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor
Classification
- Weakness
-
- CWE-400 Uncontrolled Resource Consumption
References
-
cisa.gov
/uscert/ics/advisories/icsma-22-244-01
mitigation, third-party-advisory, us-government-resource
Sources
- CVE List CVE-2022-38100
- NVD CVE-2022-38100
Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.