CVE-2022-38069 json
Contec Health CMS8000
published 2022-09-13 4 y ago modified 2026-06-17 4 mo ago also EUVD-2022-40674
- cvss
- 4.3
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.3%
- EPSS 0.3% chance of exploitation in 30 days, percentile 27%
- public exploits
- 0
- none known
Summary
Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-22-244-01 updated cvss 4.3
Contec Health CMS8000 Patient Monitor (Update A)
-
Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor <=smart3250-2.6.27-wlan2.1.7.cramfs, <=CMS7.820.075.08/0.740.75, <=CMS7.820.120.01/0.930.95affected: mitigation “Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.”
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
CISA SSVC
- Exploitation
- none
- Automatable
- no
- Technical impact
- partial
- Assessed
- 18 mo ago
Exploit prediction
0.34% chance of exploitation in the next 30 days, higher than 27% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.1 | 4.3 | CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L |
| NVD | 3.1 | 6.1 | CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N |
Affected products
- Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor
Classification
- Weakness
-
- CWE-798 Use of Hard-coded Credentials
References
-
cisa.gov
/uscert/ics/advisories/icsma-22-244-01
mitigation, third-party-advisory, us-government-resource
Sources
- CVE List CVE-2022-38069
- NVD CVE-2022-38069