CVE-2021-43355 json
Fresenius Kabi Agilia Connect Infusion System use of client side authentication
published 2022-01-21 4 y ago modified 2026-06-17 4 mo ago also EUVD-2021-30291
- cvss
- 7.3
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 1.0%
- EPSS 1.0% chance of exploitation in 30 days, percentile 62%
- public exploits
- 0
- none known
Summary
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 allows user input to be validated on the client side without authentication by the server. The server should not rely on the correctness of the data because users might not support or block JavaScript or intentionally bypass the client-side checks. An attacker with knowledge of the service user could circumvent the client-side control and login with service privileges.
Patch / fix
- Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) 2.0.1.3
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-21-355-01 updated cvss 7.3
Fresenius Kabi Agilia Connect Infusion System (Update A)
-
Fresenius Kabi Agilia Connect WiFi module of the pumps <= D25 Fresenius Kabi Agilia Link+ <= 3.0 D15 Fresenius Kabi Agilia Partner maintenance software <= 3.3.0affected: mitigation “Fresenius Kabi has created new versions to address these vulnerabilities: Link+ v3.0 (D16 or later)” fresenius-kabi.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
CISA SSVC
- Exploitation
- none
- Automatable
- yes
- Technical impact
- partial
- Assessed
- 18 mo ago
Exploit prediction
0.98% chance of exploitation in the next 30 days, higher than 62% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.1 | 7.3 | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L |
| NVD | 3.1 | 9.8 | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
| NVD | 2.0 | 7.5 | AV:N/AC:L/Au:N/C:P/I:P/A:P |
Affected products
- Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard)
Classification
References
-
cisa.gov
/uscert/ics/advisories/icsma-21-355-01
third-party-advisory, us-government-resource
Sources
- CVE List CVE-2021-43355
- NVD CVE-2021-43355
Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.