CVE-2021-38394 json
Missing Protection against Hardware Reverse Engineering Using Integrated Circuit Imaging Techniques for Boston Scientific Zoom Latitude
published 2021-10-04 5 y ago modified 2026-06-17 4 mo ago also EUVD-2021-24846
- cvss
- 6.2
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.2%
- EPSS 0.2% chance of exploitation in 30 days, percentile 14%
- public exploits
- 0
- none known
Summary
An attacker with physical access to the device can extract the binary that checks for the hardware key and reverse engineer it, which could be used to create a physical duplicate of a valid hardware key. The hardware key allows access to special settings when inserted.
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-21-273-01 updated cvss 6.2
Boston Scientific Zoom Latitude
-
Boston Scientific ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) Model 3120affected: vendor fix
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.23% chance of exploitation in the next 30 days, higher than 14% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.1 | 6.2 | CVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:C/C:L/I:H/A:L |
| NVD | 3.1 | 6.4 | CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H |
| NVD | 2.0 | 6.9 | AV:L/AC:M/Au:N/C:C/I:C/A:C |
Affected products
- Boston Scientific ZOOM LATITUDE
Classification
- Weakness
-
- CWE-1278 Missing Protection Against Hardware Reverse Engineering Using Integrated Circuit (IC) Imaging Techniques
References
-
us-cert.cisa.gov
/ics/advisories/icsma-21-273-01
third-party-advisory, us-government-resource
Sources
- CVE List CVE-2021-38394
- NVD CVE-2021-38394
Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.