An Insufficient Verification of Data Authenticity vulnerability in B.

published 2021-08-25 5 y ago modified 2026-06-17 4 mo ago also EUVD-2021-20558

cvss
10.0
version 3.1, CNA
cisa kev
Not listed
not in CISA's catalog
epss
5.6%
EPSS 5.6% chance of exploitation in 30 days, percentile 92%
public exploits
0
none known

Summary

An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets.

Patch / fix

  • bbraun spacecom2 012u000062

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-21-294-01 updated cvss 9.0

B. Braun Infusomat Space Large Volume Pump

  • B. Braun Medical Battery Pack SP with Wi-Fi <= L81 B. Braun Medical Battery pack SP with WiFi <= 028U000061 B. Braun Medical Data module compactPlus A10 < A11 B. Braun Medical SpaceStation with SpaceCom 2 <= L81, <= 012U000061
    affected: vendor fix bbraun.com

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

Exploit prediction

5.58% chance of exploitation in the next 30 days, higher than 92% of all scored CVEs (FIRST EPSS, 2026-11-08).

Exploits and checks

  • mcafee.com https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/ referenced PoC, tagged "exploit" by a source, unverified

CVSS scores

assigned by version score vector
CNA 3.1 10.0 CVSS:3.1/AC:L/AV:N/A:N/C:H/I:H/PR:N/S:C/UI:N
CNA 3.1 10.0 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:N
NVD 3.1 9.8 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
NVD 2.0 10.0 AV:N/AC:L/Au:N/C:C/I:C/A:C

Classification

Weakness
  • CWE-347 Improper Verification of Cryptographic Signature

References

Sources

Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.