CVE-2021-33885 json
An Insufficient Verification of Data Authenticity vulnerability in B.
published 2021-08-25 5 y ago modified 2026-06-17 4 mo ago also EUVD-2021-20558
- cvss
- 10.0
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 5.6%
- EPSS 5.6% chance of exploitation in 30 days, percentile 92%
- public exploits
- 0
- none known
Summary
An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets.
Patch / fix
- bbraun spacecom2 012u000062
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-21-294-01 updated cvss 9.0
B. Braun Infusomat Space Large Volume Pump
-
B. Braun Medical Battery Pack SP with Wi-Fi <= L81 B. Braun Medical Battery pack SP with WiFi <= 028U000061 B. Braun Medical Data module compactPlus A10 < A11 B. Braun Medical SpaceStation with SpaceCom 2 <= L81, <= 012U000061affected: vendor fix bbraun.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
5.58% chance of exploitation in the next 30 days, higher than 92% of all scored CVEs (FIRST EPSS, 2026-11-08).
Exploits and checks
- mcafee.com https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/ referenced PoC, tagged "exploit" by a source, unverified
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.1 | 10.0 | CVSS:3.1/AC:L/AV:N/A:N/C:H/I:H/PR:N/S:C/UI:N |
| CNA | 3.1 | 10.0 | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:N |
| NVD | 3.1 | 9.8 | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
| NVD | 2.0 | 10.0 | AV:N/AC:L/Au:N/C:C/I:C/A:C |
Classification
- Weakness
-
- CWE-347 Improper Verification of Cryptographic Signature
References
-
mcafee.com
/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-brau
exploit, third-party-advisory
-
bbraunusa.com
/en.htm
broken-link
Sources
- CVE List CVE-2021-33885
- NVD CVE-2021-33885
Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.