Skip to content
Fencer Vulns
high 7.5

CVE-2021-23236 json

Fresenius Kabi Agilia Connect Infusion System uncontrolled resource consumption

published 2022-01-21 4 y ago modified 2026-06-17 4 mo ago also EUVD-2021-10335

cvss
7.5
version 3.1, CNA
cisa kev
Not listed
not in CISA's catalog
epss
1.1%
EPSS 1.1% chance of exploitation in 30 days, percentile 65%
public exploits
0
none known

Summary

Requests may be used to interrupt the normal operation of the device. When exploited, Fresenius Kabi Agilia Link+ version 3.0 must be rebooted via a hard reset triggered by pressing a button on the rack system.

Patch / fix

  • Fresenius Kabi Agilia Link+ 3.0

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-21-355-01 updated cvss 7.5

Fresenius Kabi Agilia Connect Infusion System (Update A)

  • Fresenius Kabi Agilia Connect WiFi module of the pumps <= D25 Fresenius Kabi Agilia Link+ <= 3.0 D15 Fresenius Kabi Agilia Partner maintenance software <= 3.3.0
    affected: mitigation “Fresenius Kabi has created new versions to address these vulnerabilities: Link+ v3.0 (D16 or later)” fresenius-kabi.com

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

CISA SSVC

Exploitation
none
Automatable
yes
Technical impact
partial
Assessed
18 mo ago

Exploit prediction

1.07% chance of exploitation in the next 30 days, higher than 65% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
CNA 3.1 7.5 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
NVD 3.1 7.5 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
NVD 2.0 7.8 AV:N/AC:L/Au:N/C:N/I:N/A:C

Affected products

  • Fresenius Kabi Agilia Link+

Classification

Weakness
  • CWE-400 Uncontrolled Resource Consumption

References

Sources

Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.