In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, an XML validation vulnerability in the ventilator allows privileged attacke…
published 2021-03-15 5 y ago modified 2026-06-17 4 mo ago also EUVD-2020-19795
- cvss
- 4.3
- version 3.1, NVD
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.3%
- EPSS 0.3% chance of exploitation in 30 days, percentile 18%
- public exploits
- 0
- none known
Summary
In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, an XML validation vulnerability in the ventilator allows privileged attackers with physical access to render the device persistently unusable by uploading specially crafted configuration files.
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-21-047-01 updated cvss 4.3
Hamilton-T1
-
Hamilton Medical AG T1 Ventilator <= 2.2.3affected: vendor fix
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.26% chance of exploitation in the next 30 days, higher than 18% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| NVD | 3.1 | 4.3 | CVSS:3.1/AV:P/AC:L/PR:L/UI:N/S:U/C:N/I:N/A:H |
| NVD | 2.0 | 2.1 | AV:L/AC:L/Au:N/C:N/I:N/A:P |
Affected products
- n/a Hamilton Medical AG, T1-Ventillator
Classification
- Weakness
-
- CWE-112 Missing XML Validation
References
-
us-cert.cisa.gov
/ics/advisories/icsma-21-047-01