CVE-2020-27260 json
Innokas Yhtymä Oy Vital Signs Monitor VC150 prior to Version 1.7.15 HL7 v2.x injection vulnerabilities exist in the affected products that…
published 2021-01-08 5 y ago modified 2026-06-17 4 mo ago also EUVD-2020-19773
- cvss
- 5.3
- version 3.1, NVD
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.4%
- EPSS 0.4% chance of exploitation in 30 days, percentile 38%
- public exploits
- 0
- none known
Summary
Innokas Yhtymä Oy Vital Signs Monitor VC150 prior to Version 1.7.15 HL7 v2.x injection vulnerabilities exist in the affected products that allow physically proximate attackers with a connected barcode reader to inject HL7 v2.x segments into specific HL7 v2.x messages via multiple expected parameters.
Patch / fix
- innokasmedical vital_signs_monitor_vc150_firmware 1.7.15
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-21-007-01 updated cvss 5.3
Innokas Yhtymä Oy Vital Signs Monitor
-
Innokas Yhtymä Oy VC150 < 1.7.15affected: vendor fix
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.44% chance of exploitation in the next 30 days, higher than 38% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| NVD | 3.1 | 5.3 | CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:N |
| NVD | 2.0 | 2.1 | AV:L/AC:L/Au:N/C:N/I:P/A:N |
Affected products
- n/a BIGCOMPANYSOFT SOFTWARE PRODUCT
Classification
- Weakness
-
- CWE-74 Improper Neutralization of Special Elements in Output Used by a Downstream Component
References
-
us-cert.cisa.gov
/ics/advisories/icsma-21-007-01
third-party-advisory, us-government-resource
Sources
- CVE List CVE-2020-27260
- NVD CVE-2020-27260