high 8.8

CVE-2020-27252 json

Medtronic MyCareLink Smart Time-of-check Time-of-use Race Condition

published 2020-12-14 5 y ago modified 2026-06-17 4 mo ago also EUVD-2020-19765

cvss
8.8
version 3.1, CNA
cisa kev
Not listed
not in CISA's catalog
epss
3.8%
EPSS 3.8% chance of exploitation in 30 days, percentile 90%
public exploits
0
none known

Summary

Medtronic MyCareLink Smart 25000 is

vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-20-345-01 updated cvss 8.8

Medtronic MyCareLink Smart

  • Medtronic Smart Model 25000 Patient Reader
    affected: vendor fix medtronic.com

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

Exploit prediction

3.84% chance of exploitation in the next 30 days, higher than 90% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
CNA 3.1 8.8 CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H
NVD 3.1 8.1 CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
NVD 2.0 9.3 AV:N/AC:M/Au:N/C:C/I:C/A:C

Affected products

  • Medtronic Smart Model 25000 Patient Reader

Classification

Weakness
  • CWE-367 Time-of-check Time-of-use (TOCTOU) Race Condition

References

Sources

Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.