CVE-2020-27252 json
Medtronic MyCareLink Smart Time-of-check Time-of-use Race Condition
published 2020-12-14 5 y ago modified 2026-06-17 4 mo ago also EUVD-2020-19765
- cvss
- 8.8
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 3.8%
- EPSS 3.8% chance of exploitation in 30 days, percentile 90%
- public exploits
- 0
- none known
Summary
Medtronic MyCareLink Smart 25000 is
vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-20-345-01 updated cvss 8.8
Medtronic MyCareLink Smart
-
Medtronic Smart Model 25000 Patient Readeraffected: vendor fix medtronic.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
3.84% chance of exploitation in the next 30 days, higher than 90% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.1 | 8.8 | CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H |
| NVD | 3.1 | 8.1 | CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H |
| NVD | 2.0 | 9.3 | AV:N/AC:M/Au:N/C:C/I:C/A:C |
Affected products
- Medtronic Smart Model 25000 Patient Reader
Classification
- Weakness
-
- CWE-367 Time-of-check Time-of-use (TOCTOU) Race Condition
References
Sources
- CVE List CVE-2020-27252
- NVD CVE-2020-27252
Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.