CVE-2020-16224 json
Philips Patient Monitoring Devices Improper Handling of Length Parameter Inconsistency
published 2020-09-11 6 y ago modified 2026-06-17 4 mo ago also EUVD-2020-8190
- cvss
- 6.5
- version 3.1, NVD
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.6%
- EPSS 0.6% chance of exploitation in 30 days, percentile 48%
- public exploits
- 0
- none known
Summary
In Patient Information Center iX (PICiX) Versions C.02, C.03, the software parses a formatted message or structure but does not handle or incorrectly handles a length field that is inconsistent with the actual length of the associated data, causing the application on the surveillance station to restart.
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-20-254-01 updated cvss 6.5
Philips Patient Monitoring Devices (Update C)
-
Philips IntelliVue patient monitor MP2-MP90 <=N Philips IntelliVue patient monitor MX100 <=N Philips IntelliVue patient monitor MX400-MX850 <=Naffected: vendor fix, mitigation “Philips released the following versions to remediate reported vulnerabilities:” usa.philips.com
-
Philips IntelliVue X2 <=N Philips IntelliVue X3 <=Naffected: mitigation “Philips released the following versions to remediate reported vulnerabilities:” incenter.medical.philips.com
-
Philips Patient Information Center iX (PICiX) B.02, C.02, C.03 Philips PerformanceBridge Focal Point A.01affected: vendor fix, mitigation “Philips released the following versions to remediate reported vulnerabilities:” incenter.medical.philips.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.59% chance of exploitation in the next 30 days, higher than 48% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| NVD | 3.1 | 6.5 | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H |
| NVD | 2.0 | 3.3 | AV:A/AC:L/Au:N/C:N/I:N/A:P |
Affected products
- Philips Patient Information Center iX (PICiX)
Classification
- Weakness
-
- CWE-130 Improper Handling of Length Parameter Inconsistency
References
-
us-cert.cisa.gov
/ics/advisories/icsma-20-254-01
third-party-advisory, us-government-resource
- philips.com /productsecurity
Sources
- CVE List CVE-2020-16224
- NVD CVE-2020-16224
Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.