CVE-2020-15486 json
An issue was discovered on Dr Trust ECG Pen 2.00.08 devices.
published 2020-08-26 6 y ago modified 2026-06-17 4 mo ago also EUVD-2020-7479
- cvss
- 6.5
- version 3.1, NVD
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.5%
- EPSS 0.5% chance of exploitation in 30 days, percentile 43%
- public exploits
- 0
- none known
Summary
An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.
Exploit prediction
0.51% chance of exploitation in the next 30 days, higher than 43% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| NVD | 3.1 | 6.5 | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N |
| NVD | 2.0 | 3.3 | AV:A/AC:L/Au:N/C:P/I:N/A:N |