CVE-2020-12048 json
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., T…
published 2020-06-29 6 y ago modified 2026-06-17 4 mo ago also EUVD-2020-4364
- cvss
- 7.5
- version 3.1, NVD
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.5%
- EPSS 0.5% chance of exploitation in 30 days, percentile 39%
- public exploits
- 0
- none known
Summary
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-20-170-03 updated cvss 7.5
Baxter Phoenix Hemodialysis Delivery System (Update A)
-
Baxter Phoenix Hemodialysis Delivery System 3.36, 3.40affected: vendor fix baxter.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.45% chance of exploitation in the next 30 days, higher than 39% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| NVD | 3.1 | 7.5 | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N |
| NVD | 2.0 | 5.0 | AV:N/AC:L/Au:N/C:P/I:N/A:N |
Affected products
- n/a Baxter Phoenix Hemodialysis Delivery System
Classification
- Weakness
-
- CWE-319 Cleartext Transmission of Sensitive Information
References
-
us-cert.gov