Skip to content
Fencer Vulns
critical 9.8

CVE-2020-12040 json

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the applicat…

published 2020-06-29 6 y ago modified 2026-06-17 4 mo ago also EUVD-2020-4356

cvss
9.8
version 3.1, NVD
cisa kev
Not listed
not in CISA's catalog
epss
0.9%
EPSS 0.9% chance of exploitation in 30 days, percentile 61%
public exploits
0
none known

Summary

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-20-170-04 updated cvss 7.3

Baxter Sigma Spectrum Infusion Pumps (Update B)

  • Baxter Spectrum v8.x model 35700BAX2, v9.x model 35700BAX3 Baxter Spectrum LVP v8.x (with Wireless Battery Modules v17, v17D19, >= v20D29, <= v20D32, >= v22D24, <= v22D28), v9.x (with Wireless Battery Module >= v22D19, <= v22D28) Baxter Sigma Spectrum v6.x model 35700BAX Baxter Sigma Spectrum LVP v6.x (with Wireless Battery Modules v9, v11, v13, v14, v15, v16, v16D38, v17, v17D19, >= v20D29, <=
    affected: vendor fix baxter.com

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

Exploit prediction

0.95% chance of exploitation in the next 30 days, higher than 61% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
NVD 3.1 9.8 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
NVD 2.0 5.0 AV:N/AC:L/Au:N/C:P/I:N/A:N

Affected products

  • n/a Baxter Sigma Spectrum Infusion Pumps

Classification

Weakness
  • CWE-319 Cleartext Transmission of Sensitive Information

References

Sources

Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.