CVE-2020-12040 json
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the applicat…
published 2020-06-29 6 y ago modified 2026-06-17 4 mo ago also EUVD-2020-4356
- cvss
- 9.8
- version 3.1, NVD
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.9%
- EPSS 0.9% chance of exploitation in 30 days, percentile 61%
- public exploits
- 0
- none known
Summary
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-20-170-04 updated cvss 7.3
Baxter Sigma Spectrum Infusion Pumps (Update B)
-
Baxter Spectrum v8.x model 35700BAX2, v9.x model 35700BAX3 Baxter Spectrum LVP v8.x (with Wireless Battery Modules v17, v17D19, >= v20D29, <= v20D32, >= v22D24, <= v22D28), v9.x (with Wireless Battery Module >= v22D19, <= v22D28) Baxter Sigma Spectrum v6.x model 35700BAX Baxter Sigma Spectrum LVP v6.x (with Wireless Battery Modules v9, v11, v13, v14, v15, v16, v16D38, v17, v17D19, >= v20D29, <=affected: vendor fix baxter.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.95% chance of exploitation in the next 30 days, higher than 61% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| NVD | 3.1 | 9.8 | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
| NVD | 2.0 | 5.0 | AV:N/AC:L/Au:N/C:P/I:N/A:N |
Affected products
- n/a Baxter Sigma Spectrum Infusion Pumps
Classification
- Weakness
-
- CWE-319 Cleartext Transmission of Sensitive Information
References
-
us-cert.gov
/ics/advisories/icsma-20-170-04
third-party-advisory, us-government-resource
Sources
- CVE List CVE-2020-12040
- NVD CVE-2020-12040