Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35…

published 2020-06-29 6 y ago modified 2026-06-17 4 mo ago also EUVD-2020-4355

cvss
2.4
version 3.1, NVD
cisa kev
Not listed
not in CISA's catalog
epss
0.3%
EPSS 0.3% chance of exploitation in 30 days, percentile 28%
public exploits
0
none known

Summary

Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-20-170-04 updated cvss 4.3

Baxter Sigma Spectrum Infusion Pumps (Update B)

  • Baxter Spectrum v8.x model 35700BAX2, v9.x model 35700BAX3 Baxter Spectrum LVP v8.x (with Wireless Battery Modules v17, v17D19, >= v20D29, <= v20D32, >= v22D24, <= v22D28), v9.x (with Wireless Battery Module >= v22D19, <= v22D28) Baxter Sigma Spectrum v6.x model 35700BAX Baxter Sigma Spectrum LVP v6.x (with Wireless Battery Modules v9, v11, v13, v14, v15, v16, v16D38, v17, v17D19, >= v20D29, <=
    affected: vendor fix baxter.com

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

Exploit prediction

0.35% chance of exploitation in the next 30 days, higher than 28% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
NVD 3.1 2.4 CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N
NVD 2.0 2.1 AV:L/AC:L/Au:N/C:P/I:N/A:N

Affected products

  • n/a Baxter Sigma Spectrum Infusion Pumps

Classification

Weakness
  • CWE-259 Use of Hard-coded Password
  • CWE-798 Use of Hard-coded Credentials

References