CVE-2019-5024 json
A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical inform…
published 2019-04-11 7 y ago modified 2026-06-17 4 mo ago also EUVD-2019-14631
- cvss
- 7.6
- version 3.0, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.5%
- EPSS 0.5% chance of exploitation in 30 days, percentile 41%
- public exploits
- 0
- none known
Summary
A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical information collection devices running versions 9.0.3 or lower. A specific series of keyboard inputs can escape the restricted environment, resulting in full administrator access to the underlying operating system. An attacker can connect to the device via USB port with a keyboard or other HID device to trigger this vulnerability.
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-20-196-01 updated cvss 7.6
Capsule Technologies SmartLinx Neuron 2 (Update A)
-
Capsule Technologies SmartLinx Neuron 2 <= 9.0.3affected: vendor fix
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.47% chance of exploitation in the next 30 days, higher than 41% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.0 | 7.6 | CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
| NVD | 3.1 | 7.6 | CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
| NVD | 2.0 | 7.2 | AV:L/AC:L/Au:N/C:C/I:C/A:C |
Affected products
- Capsule Technologies Capsule Technologies SmartLinx Neuron 2
Classification
- Weakness
-
- CWE-693 Protection Mechanism Failure
References
-
talosintelligence.com
/vulnerability_reports/TALOS-2019-0785
third-party-advisory
Sources
- CVE List CVE-2019-5024
- NVD CVE-2019-5024