CVE-2019-25724 json
Dräger Infinity M300 VG2.x Network-Based Denial of Service
published 2026-06-02 5 mo ago modified 2026-07-22 3 mo ago also EUVD-2019-20160
- cvss
- 6.5
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.2%
- EPSS 0.2% chance of exploitation in 30 days, percentile 6%
- public exploits
- 0
- none known
Summary
Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.
Patch / fix
- Dräger Infinity M300 VG3.0
CISA SSVC
- Exploitation
- none
- Automatable
- no
- Technical impact
- partial
- Assessed
- 5 mo ago
Exploit prediction
0.18% chance of exploitation in the next 30 days, higher than 6% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 4.0 | 7.1 | CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N |
| CNA | 3.1 | 6.5 | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H |
| CNA | 4.0 | 7.1 | CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X |
Affected products
- Dräger Infinity M300
Classification
- Weakness
-
- CWE-400 Uncontrolled Resource Consumption
References
Sources
- CVE List CVE-2019-25724
- NVD CVE-2019-25724