CVE-2019-25723 json

Dräger Perseus A500 2.00-2.02 DoS via Medibus Interface

published 2026-06-02 5 mo ago modified 2026-07-22 3 mo ago also EUVD-2019-20159

cvss
4.0
version 3.1, CNA
cisa kev
Not listed
not in CISA's catalog
epss
0.2%
EPSS 0.2% chance of exploitation in 30 days, percentile 14%
public exploits
0
none known

Summary

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

Patch / fix

  • Dräger Perseus A500 2.03

CISA SSVC

Exploitation
none
Automatable
no
Technical impact
partial
Assessed
5 mo ago

Exploit prediction

0.24% chance of exploitation in the next 30 days, higher than 14% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
CNA 4.0 6.3 CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L
CNA 3.1 4.0 CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L
CNA 4.0 6.3 CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

Affected products

  • Dräger Perseus A500

Classification

Weakness
  • CWE-1286 Improper Validation of Syntactic Correctness of Input

References