CVE-2019-18256 json
BIOTRONIK CardioMessenger II, The affected products use individual per-device credentials that are stored in a recoverable format.
published 2020-06-29 6 y ago modified 2026-06-17 4 mo ago also EUVD-2019-8054
- cvss
- 4.6
- version 3.1, NVD
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.4%
- EPSS 0.4% chance of exploitation in 30 days, percentile 29%
- public exploits
- 0
- none known
Summary
BIOTRONIK CardioMessenger II, The affected products use individual per-device credentials that are stored in a recoverable format. An attacker with physical access to the CardioMessenger can use these credentials for network authentication and decryption of local data in transit.
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-20-170-05 updated cvss 4.6
BIOTRONIK CardioMessenger II
-
BIOTRONIK CardioMessenger II-S GSM T4APP 2.2 BIOTRONIK CardioMessenger II-S T-Line T4APP 2.2affected: vendor fix
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.36% chance of exploitation in the next 30 days, higher than 29% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| NVD | 3.1 | 4.6 | CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N |
| NVD | 2.0 | 2.1 | AV:L/AC:L/Au:N/C:P/I:N/A:N |
Affected products
- n/a BIOTRONIK CardioMessenger II-S T-Line, CardioMessenger II-S GSM
Classification
References
-
us-cert.gov
/ics/advisories/icsma-20-170-05