Medtronic MiniMed 508 and Paradigm Series Insulin Pumps Improper Access Control

published 2019-06-28 7 y ago modified 2026-06-17 4 mo ago also EUVD-2019-2678

cvss
7.1
version 3.1, CNA
cisa kev
Not listed
not in CISA's catalog
epss
1.2%
EPSS 1.2% chance of exploitation in 30 days, percentile 69%
public exploits
0
none known

Summary

Medtronic MiniMed Insulin Pumps

are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-19-178-01 updated cvss 7.1

Medtronic MiniMed 508 and Paradigm Series Insulin Pumps

  • Medtronic MiniMed 508 pump Medtronic MiniMed Paradigm 511 pump Medtronic MiniMed Paradigm 512/712 pumps Medtronic MiniMed Paradigm 515/715 pumps Medtronic MiniMed Paradigm 522/722 pumps Medtronic MiniMed Paradigm 522K/722K pumps
    +5 more products
    Medtronic MiniMed Paradigm 523/723 pumps <= 2.4A Medtronic MiniMed Paradigm 523K/723K pumps <= 2.4A Medtronic MiniMed Paradigm 712E pump Medtronic MiniMed Paradigm Veo 554/754 pumps <= 2.6A Medtronic MiniMed Paradigm Veo 554CM and 754CM models only <= 2.7A
    affected: vendor fix medtronic.com

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

Exploit prediction

1.23% chance of exploitation in the next 30 days, higher than 69% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
CNA 3.1 7.1 CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H
NVD 3.0 8.8 CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
NVD 2.0 5.8 AV:A/AC:L/Au:N/C:P/I:P/A:P

Affected products

  • Medtronic MiniMed 508 pump
  • Medtronic MiniMed Paradigm 511 pump
  • Medtronic MiniMed Paradigm 512/712 pumps
  • Medtronic MiniMed Paradigm 712E pump
  • Medtronic MiniMed Paradigm 515/715 pumps
  • Medtronic MiniMed Paradigm 522/722 pumps
  • Medtronic MiniMed Paradigm 522K/722K pumps
  • Medtronic MiniMed Paradigm 523/723 pumps
  • Medtronic MiniMed Paradigm 523K/723K pumps
  • Medtronic MiniMed Paradigm Veo 554/754 pumps
  • Medtronic MiniMed Paradigm Veo 554CM/754CM pumps