CVE-2018-4845 json
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use o…
published 2018-06-26 8 y ago modified 2026-06-17 4 mo ago also EUVD-2018-16631
- cvss
- 8.8
- version 3.1, NVD
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 1.2%
- EPSS 1.2% chance of exploitation in 30 days, percentile 67%
- public exploits
- 0
- none known
Summary
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). Remote attackers with either local or remote credentialed access to the "Remote View" feature might be able to elevate their privileges, compromising confidentiality, integrity, and availability of the system. No special skills or user interaction are required to perform this attack. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.
Patch / fix
- siemens rapidlab_1200_firmware 3.3
Exploit prediction
1.16% chance of exploitation in the next 30 days, higher than 67% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| NVD | 3.1 | 8.8 | CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H |
| NVD | 2.0 | 6.5 | AV:N/AC:L/Au:S/C:P/I:P/A:P |
Affected products
- Siemens AG RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems, RAPIDLab 1200 Series, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 400 systems
Classification
References
-
cert-portal.siemens.com
/productcert/pdf/ssa-755010.pdf
third-party-advisory
Sources
- CVE List CVE-2018-4845
- NVD CVE-2018-4845
Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.