CVE-2018-14781 json
Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay
published 2018-08-13 8 y ago modified 2026-06-17 4 mo ago also EUVD-2018-6665
- cvss
- 5.3
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.7%
- EPSS 0.7% chance of exploitation in 30 days, percentile 53%
- public exploits
- 0
- none known
Summary
Medtronic MiniMed MMT
devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-18-219-02 updated cvss 5.3
Medtronic MiniMed MMT-500/MMT-503 Remote Controllers (Update A)
-
Medtronic MMT-500 Remote Controller MMT-508 MiniMed pump Medtronic MMT-503 Remote Controller MMT-512 / MMT-712 Paradigm x12, MMT-523 / MMT-723 Paradigm Revel, MMT-523(K) / MMT-723(K) Paradigm, MMT-515 / MMT-715 Paradigm x15affected: vendor fix medtronic.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.71% chance of exploitation in the next 30 days, higher than 53% of all scored CVEs (FIRST EPSS, 2026-11-08).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.1 | 5.3 | CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N |
| NVD | 3.0 | 5.3 | CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N |
| NVD | 2.0 | 2.9 | AV:A/AC:M/Au:N/C:N/I:P/A:N |
Affected products
- Medtronic MMT- 508 - MiniMed pump
- Medtronic MMT – 511 pump Paradigm
- Medtronic MMT – 512 / MMT – 712 Paradigm x12
- Medtronic MMT – 515 / MMT – 715 Paradigm x15
- Medtronic MMT – 522 / MMT – 722 Paradigm REAL-TIME
- Medtronic MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME
- Medtronic MMT – 523 / MMT – 723 Paradigm Revel
- Medtronic MMT – 523(K) / MMT – 723(K) Paradigm
- Medtronic MMT – 554 / MMT – 754 MiniMed Veo
- Medtronic MMT – 551 / MMT – 751 MiniMed 530G
Classification
References
- global.medtronic.com /xg-en/product-security/security-bulletins/minimed.html
-
ics-cert.us-cert.gov
/advisories/ICSMA-18-219-02
third-party-advisory, us-government-resource
-
securityfocus.com
/bid/105044
vdb-entry, third-party-advisory
Sources
- CVE List CVE-2018-14781
- NVD CVE-2018-14781
Fencer Vulns keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.