Skip to content
Fencer Vulns
high 7.5

CVE-2016-5085 json

Johnson & Johnson Animas OneTouch Ping devices do not properly generate random numbers, which makes it easier for remote attackers to spoof…

published 2016-10-05 10 y ago modified 2026-06-17 4 mo ago also EUVD-2016-6038

cvss
7.5
version 3.0, NVD
cisa kev
Not listed
not in CISA's catalog
epss
3.9%
EPSS 3.9% chance of exploitation in 30 days, percentile 90%
public exploits
0
none known

Summary

Johnson & Johnson Animas OneTouch Ping devices do not properly generate random numbers, which makes it easier for remote attackers to spoof meters by sniffing the network and then engaging in an authentication handshake.

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-16-279-01 updated cvss 4.2

Animas OneTouch Ping Insulin Pump Vulnerabilities

  • Animas OneTouch Ping insulin pump system
    affected: mitigation “Animas has provided the following compensating controls to help reduce the risk associated with the exploitation of the identified vulnerabilities:” animas.com

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

Exploit prediction

3.86% chance of exploitation in the next 30 days, higher than 90% of all scored CVEs (FIRST EPSS, 2026-11-08).

CVSS scores

assigned by version score vector
NVD 3.0 7.5 CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
NVD 2.0 7.8 AV:N/AC:L/Au:N/C:N/I:C/A:N

Classification

Weakness
  • CWE-330 Use of Insufficiently Random Values

References

Sources